X. Chen, L. Wang, P. Fallavollita, N. Navab
Error analysis of the x-ray projection geometry of camera-augmented mobile C-arm SPIE Medical Imaging, San Diego, California, USA, February 2012 (bib) |
||
The Camera-Augmented Mobile C-arm (CamC) augments X-ray by camera images and is calibrated in a suboptimal way for X-ray to be warped onto optical images in order to align them and have common projection geometry. We investigated and compared calibration and distortion correction accuracy of: (i) the existing CAMC method (ii) Zhang's method for optical cameras, and (iii) traditional C-arm method. The RMS errors based on a synthetic model and pose estimation shows that the traditional C-arm method (µ=0.39 pixels) outperforms Zhang (µ=0.68 pixels) and CAMC (µ=1.07 pixels) methods. | ||
This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each authors copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder. |