PublicationDetail

Chair for Computer Aided Medical Procedures & Augmented Reality
Lehrstuhl für Informatikanwendungen in der Medizin & Augmented Reality

THIS WEBPAGE IS DEPRECATED - please visit our new website

A. Martinez-Möller, M. Souvatzoglou, S. Nekolla
Photon attenuation correction in whole-body PET/MRI using tissue classification
Proceedings of IEEE International Symposium on Biomedical Imaging: From Nano to Macro (ISBI 2010), pp. 1061-1062, Rotterdam, Netherlands, April 2010 - invited talk (bib)

A challenge for hybrid PET/MRI imaging is the determination of the photon attenuation map to correct the acquired PET data. Segmentation of adapted MRI sequence data in four tissue classes (background, lungs, fat, and soft tissue) is proposed and evaluated using data from 35 PET/CT whole-body oncological examinations as well as two PET/CT and MRI examinations from the same patients. When reconstructing with the segmented attenuation map, small differences in tumor uptake were observed mainly for osseous lesions, but did not change the clinical interpretation in any case. The method appears to be viable for clinical use.
This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each authors copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.