T. Cheslerean-Boghiu, F. Pfeiffer, T. Lasser
Task-driven Acquisition in Anisotropic X-ray Dark-Field Tomography
International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine (Fully3D), Philadelphia, USA, June 2019 (bib)

Anisotropic X-ray Dark-field Tomography (AXDT) is a novel imaging modality aimed at the reconstruction of spherical scattering functions in every three-dimensional volume element, based on the directional X-ray dark-field contrast as measured by an X-ray grating interferometer. In this work, we re-derive a detectability index for the AXDT forward model directly using the spherical function formulation, and use it to compute optimized acquisition trajectories using a greedy algorithm. The results demonstrate that the optimized trajectories can represent task-specific features in AXDT accurately using only a fraction of the data.
This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each authors copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.