W. Wein, A. Ladikos, A. Baumgartner
Self-calibration of geometric and radiometric parameters for cone-beam computed tomography
11th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, July 11-15 2011, Potsdam, Germany. (bib)

Thanks to the advances in parallel processing hardware, iterative algorithms for cone beam reconstruction are now available with computation times acceptable for clinical use. At the same time they are able to accomodate more accurately the physical effects underlying the X-Ray imaging process. Many parameters are involved, which need to be precisely calibrated in order to achieve an accurate 3D reconstruction. Unfortunately, some parameters might change individually for every cone beam acquisition, stirring the need for an online calibration technique. We present a method for automatically deriving individual parameter adjustements within an iterative reconstruction framework, without the need for a designated calibration phantom. Preliminary results on a beads phantom and anatomical sample show that self-calibration of global and local geometric parameters is possible; besides we briefly demonstrate radiometric calibration on phantom data.
This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each authors copyright. In most cases, these works may not be reposted without the explicit permission of the copyright holder.



Edit | Attach | Refresh | Diffs | More | Revision r1.13 - 30 Jan 2019 - 15:16 - LeslieCasas

Lehrstuhl für Computer Aided Medical Procedures & Augmented Reality    rss.gif